Comparison of upper limb muscle activity in four walking canes: A preliminary study

Faye Y. Chiou-Tan, MD; Kevin N. Magee, PhD; Thomas A. Krouskop, PhD

Department of Physical Medicine and Rehabilitation and the Division of Bioengineering, Baylor College of Medicine, Houston, TX 77004

Abstract--It is hypothesized that the cane and forearm can be aligned to reduce muscular activity needed during ambulation. In this prospective study, 10 nonimpaired control subjects were tested both in stationary and dynamic weight-bearing states while using different cane prototypes and the platform cane (PC) as compared to the standard cane (SC). The outcome measures were: 1) root mean square (RMS) voltage (µV) of electromyographic (EMG) signal as a measure of muscle power and 2) distance of ambulation. Results of stationary cane use showed that Prototype 1 decreased RMS output by 19 percent (p=0.01), Prototype 2 with wrist splint decreased it by 23 percent (p=0.003), and the PC decreased it by 68 percent (p<0.0001) as compared to the SC (ANOVA, posthoc LSM). In conclusion, the two prototypes and the PC significantly decrease RMS voltage muscle output in the upper limb, compared to the SC.

Key words: biomechanics, cane, gait aid, upper limb.


This material is based upon work supported by the Estate of Lucille Hansen and the Baylor College of Medicine Young Faculty Grant.
Address all correspondence and requests for reprints to: Dr. Faye Chiou Tan, Department of Physical Medicine and Rehabilitation, Quentin Mease Hospital, 3601 N. MacGregor Way, Suite 202, Houston, TX 77004.

INTRODUCTION

  There has been an increase in the number of patients diagnosed with musculotendinous, neuropathic, or arthritic conditions of the arms as a result of repetitive trauma. Walking canes have also been implicated in worsening these conditions. Thus far, literature on use of the cane has focused on decreasing the muscle exertion in the leg but not the arm (1-5).

  The upper limb has not been well studied as a weight-bearing limb. In evolutionary terms, clawed quadripeds bore weight through the shaft of the long bones that served as pylons. The hand or foot was used simply as a springboard and balance device. Similarly, as infants, we crawl on our hands with our wrists extended, but always bear weight predominantly through the wrist rather than the hand. Gorillas bear weight through their knuckles. This assists in maintaining a neutral wrist alignment and allows continued weight bearing through the long-bone axis. In contrast, the human who uses a cane hyperextends his wrist when weight bearing, causing a malaligned vector of force through the hand at the level of the metacarpal bones. This creates torque about the wrist joint. Any person who has temporarily required the use of a cane or crutch knows the discomfort involved. It is well documented that chronic cane or crutch use leads to repetitive stress disorders such as tendinitis, carpal tunnel syndrome, and osteoarthritis (6-8). Unfortunately, many people with arthritis are already at risk for these problems due to their inflamed joints (9-11). For patients with carpal tunnel syndrome, Gelberman, et al. has shown that hyperflexion or extension of the wrist can lead to increased intracarpal wrist pressures (12).

  The basic design of the cane has not changed over time. People with arthritis continue to be issued crook-shaped aluminum or wooden canes, with their only adaptation being height adjustment. For the occasional cane user, this may be adequate; however, for a person with painful hand and wrist problems, daily use of a simple cane over several years to decades may be deleterious. Consequently, patients may opt to sit in a wheelchair, which leads to further physical deconditioning and joint contractures and may preclude future attempts to regain the ability to walk, decreasing the person's access to environment and work. It is the goal of this project to help these patients avoid this spiral of disability through a redesign of the cane and the forearm crutch.

  The specific hypothesis is that a prototype cane can be made to decrease muscle voltage output in the upper limb during static and dynamic weight bearing.

 

METHODS

Experimental Design
  The project was a prospective study comparing the standard cane (SC) to the platform cane (PC) and two cane prototypes, in which each of the 10 nonimpaired subjects served as his/her own control. Inclusion criteria were healthy male and female subjects aged 18 to 65; exclusion criteria were cardiopulmonary and neurocognitive disorders. Informed consent was obtained according to the rules of the local institutional review board and OSHA regulations. Confidentiality was maintained throughout the study.

Protocol
  Stationary Measurements of Cane Prototypes   All canes were adjusted to the height of the individual; that is, from floor to femoral head (13). The subject was asked to weight bear maximally on the cane with his right arm against a force gauge (Newtons). Seventy-five percent of maximal force was used as the subject's individual load. This load was constant throughout the testing of the different canes. The foot-flat stage of gait was used for this portion of the study.

  Metal surface disc EMG electrodes (Ag/AgCl, 1 cm) were coated with electrode gel and taped to the skin overlying the following muscles: infraspinatus (IS), rhomboid, latissimus dorsi (LD), biceps, triceps, deltoid, brachioradialis (BR), extensor carpi radialis (ECR), flexor carpi radialis (FCR), flexor digitorum profundus (FDP), flexor carpi ulnaris (FCU), and flexor digitorum superficialis (FDS). The active electrode was placed over the midpoint of the muscle belly and the reference electrode over the tendon. A 3-cm ground electrode was placed over the adjacent bony prominence.

  The subject was then asked to bear weight for 4 s at 75 percent maximum load against a force gauge for each different cane tested. These included: the SC, Prototype 1 (P1: weight bearing on the heel of the hand), Prototype 2 (P2: weight bearing on the knuckles), and the PC. The sequence of the canes presented was randomized. Diagrams of these prototypes are shown in Figure 1. The first three of these were evaluated with and without wrist splints to determine whether this accessory were needed.

Prototypes

Figure 1.
Illustration of prototype canes used in this study
.

  The surface EMG signal was acquisitioned with a Viking IVD EMG instrument. The root mean square (RMS) value was measured in µV for each cane in a stationary postition. The RMS is accepted as a reflection of the muscle power (14). Settings for the EMG machine were a sweep speed of 20 msec/division and gain setting of 500 µV/division. The filter band pass was set at 10-10,000 Hz. A 400 ms sample of EMG activity with average RMS voltage output (µV) was recorded to floppy disc and printed out for data analysis.

Kinetic Measurements of Cane Prototypes
  To evaluate the effect of gait on the different cane prototypes, the subject was wired with telemetric surface EMG electrodes. Disposable electro-cardiogram surface electrodes with a 1-cm active recording area surrounded by a 1-cm adhesive ring (Blue Sensor, M-00-s, Medicotest, Olstykke, Denmark) were used for active, reference, and ground recordings. The electrodes were placed in active and reference pairs 4 cm apart on the following muscles: IS, LD, deltoid, biceps, triceps, ECR, FCR, FCU, and FDS. The ground electrode was placed on the elbow; all were connected to a waist pack that sent the telemetric signal to a receiving antenna on a preamplifier, after which it was processed by the telemetric EMG system provided by Noraxon, Inc. The voltage output of these eight muscles was displayed over eight individual channels on the EMG machine. Instrumentation was as described previously, except the gain setting was 1,000 µV/division. The sweep speed was 100 pt/s. With this sweep speed, up to 14 full gait cycles can be recorded with each sampling.

  The subject's left leg was placed in a knee-immobilizing brace to alter normal gait pattern, it having been previously shown by Diaz, et al. that use of a knee brace alters the EMG activity in the leg muscles (15). The subject was then tested with the SC, PC, P1, and P2 with wrist splint. The sequence of the canes tested was randomized. The subject's distance of ambulation after walking for 1 min was measured in feet. There was a 10-min rest period between ambulation with each cane to avoid fatigue.

Data Analysis
  Statistical analysis for the experiments was performed using ANOVA with post-hoc least square mean (LSM) analysis. If the expected difference in measurement was 25 percent reduction in electromyogram RMS muscle voltage, a power analysis indicates that with alpha=0.05, beta=0.20, and power=0.80 or greater, n=10 subjects would needed for each experiment.

 

RESULTS

Static Measurements
  There were four men and six women in the study. The average height (±SD) was 66(±3) in, average weight was 139(±31) lb. The age ranged from 24 to 46 yr, with a mean of 35(±8) yr.

  Results showed a significant decrease in upper limb RMS output from static use, with and without wrist splint, of the SC (194 and 195 µV, respectively) to P1 (157 and 163 µV, respectively), the P2 (162 and 150 µV, respectively), and PC (64 µV: Table 1, Figure 2). Of the individual muscles examined, the biceps (p=0.004) and latissimus dorsi (p=0.026) muscles showed significant changes on the one-way ANOVA, on which several muscles showed borderline changes (Table 2).


Table 1.
Stationary surface electromyographic voltage recordings in weight-bearing upper limbs using walking canes.
Cane RMS (µV±SE)* p value**

SC 194±14 --
SC*** 157±14 0.935
P1 157±14 0.010
P1*** 163±14 0.033
P2 162±14 0.027
P2*** 150±14 0.003
PC  64±14 <0.0001

* Mean of all muscles tested; ** statistical analysis performed with one-way ANOVA (<0.0001) with post hoc pairwise comparison with Least Square Means (LSM); *** with wrist splint; SC=standard cane; P1=prototype 1; P2=prototype 2; PC=platform cane; values shown are in reference to the standard cane.
 

Table data
Figure 2.
Graph of Table 1 displaying average stationary surface EMG voltage muscle output for the upper limb for each type of cane used in control subjects. * Indicates p<0.05, ANOVA with post hoc LSM analysis. Error bars indicate SE.


Table 2.
Stationary surface EMG voltage recordings in individual
muscles using walking canes.
Muscle p value* Cane RMS (µV) SC/Listed p value**

Biceps 0.004 PC 198/53 0.0001
LD 0.026 PC 179/97 0.001 
BR 0.060 PC 301/64 0.002 
    P2  301/160 0.05  
    P2***  301/133 0.021 
    P1***  301/157 0.053 
Deltoid 0.330 PC 224/60 0.020 
ECRB/L 0.092 PC 186/58 0.012 
FCR 0.125 PC 161/50 0.009 
FCU 0.184 PC 162/72 0.029 
FDP 0.092 PC 178/70 0.007 
FDS 0.082 PC 153/52 0.013 
IS 0.481 PC 135/62 0.105 
Rhomboid 0.268 PC 135/45 0.043 
Triceps 0.075 PC 314/77 0.004 

* ANOVA; ** pairwise comparison to standard cane; *** with wrist splint; SC=standard cane; P1=prototype 1; P2=prototype 2; PC=platform cane; LD=latissimus dorsi; BR=brachioradialis; ECRB/L=extensor carpi radialis brevis and longus; FCR=flexor carpi radialis; FCU=flexor carpi ulnaris; FDP=flexor digitorum profundus; FDS=flexor digitorum superficialis; IS=infraspinatus.

  In this comparison, there was no significant difference in any of the canes with and without the use of wrist splints (Table 3).


Table 3.
Comparison of EMG voltage in subjects using canes with wrist splints versus using canes without.
Cane p-value*

SC 0.935
P1 0.674
P2 0.419

* Statistical analysis performed with one-way ANOVA with post hoc pairwise comparison with Least Square Means; SC=standard cane; P1=prototype 1; P2=prototype 2.

  As a result of this first study, it was concluded that P1, P2, and PC would be compared to the SC in the second part of the experiment.

Dynamic Measurements
  Results of the second study showed that during ambulation there was a significant decrease in upper limb muscle voltage when using the PC (59 µV) as compared to the SC (79 µV: Table 4, Figure 3). There was no significant difference in the efficiency of gait among the four types tested (Table 5).


Table 4.
Dynamic telemetric surface EMG voltage recordings in weight-bearing upper limbs using walking canes.
Cane RMS (µV±SE)* p value**
SC 79±55)*  -- 
P1 66±40)* 0.065
P2/w wrist splint 73±51)* 0.40 
PC 59±40)* 0.005

* Mean of all muscles tested; ** statistical analysis performed with one way ANOVA (p=0.031) with post hoc pairwise comparison with Least Square Means. SC=standard cane; P1=prototype 1; P2=prototype 2; PC=platform cane; values shown are in reference to the standard cane.

Voltage
Figure 3.
Graph of Table 4 displaying average dynamic telemetric surface EMG voltage output in upper limbs using different walking canes in control subjects. * Indicates p<0.05, ANOVA with post hoc LSM analysis. Error bars indicate SE.


Table 5.
Distance of ambulation during dynamic telemetric cane measurements.

Cane Distance (ft±SE)

SC 55±26)
P1 52±29)
P2 53±30)
PC 54±31)

Statistical analysis performed with ANOVA, repeated measures (p=0.619); SC=standard cane; P1=prototype 1; P2=prototype 2; PC=platform cane.

 

DISCUSSION

  In this study, it was found that the three canes tested were more effective in decreasing the voltage output of muscles in the upper limb than was the SC during stationary weight bearing (Table 1). Therefore, we conclude that all three canes are an improvement over the SC with regard to muscle voltage output. It was interesting to note that the improvement occurred with proximal and distal muscles tested when using the PC, and with the brachioradialis in all three canes (Table 2). Although one would expect the PC to decrease energy output distally where the arm is immobilized, and increase proximally where muscle substitution would occur, this was not found to be the case. One possibility is that the upper fibers of the trapezius muscle may be responsible for the increased effort, but this muscle was unfortunately not included in the study. A second observation is that the BR muscle activity was decreased in all three canes relative to the SC. This is likely due to the fact the wrist-cane junction was stabilized in all three canes, thereby decreasing the work by the BR in swing through, and possibly increasing the work on the anterior deltoid muscle.

  It is interesting to note that there was no significant difference in use of any canes with or without wrist splints (Table 3). This suggests that the lessened muscle activities were derived, not from stabilization of the wrist joint, but of the lessened demand for stabilizing the wrist/hand to the cane. It is also important to note that none of the three canes tested decrease the efficiency of gait relative to the SC (Table 5). This is in contrast to the subjective perception of many patients, who feel that the PC impairs their speed of gait.

  Finally, it is important to note that the PC decreases muscle voltage output dynamically as well as statically. This may be reflected in the long run by increased patient endurance and therefore enhanced level of function as a result. This would need to be tested with patients in the future, however. It is important to note that the prototype canes do not increase the amount of muscle voltage more than the SC dynamically.

Limitations of Study
  The limitations of the study are use of surface electrodes, possible electrode movement, training effect, and individual variability. While their use limits the representation of deep muscles, we employed surface electrodes for both portions of the study for greater consistency: fine wire electrodes were not available for dynamic recordings with the Noraxon system.

  The possibility of electrode movement was minimized in the static portion of the test by limiting the movement of the subject and by fixing the position of the electrode while changing canes. In the dynamic portion of the study, snap-on electrodes similar to those found in Holter monitoring were used to allow some electrode movement without affecting the recording.

  The training effect was minimized by asking each subject to practice using the cane until a level of comfort had been reached. The time it took to learn how to use each cane varied from individual to individual and cane to cane. Each training session was followed by a rest period of 10 min before testing began.

  Individual variability from differences in size and strength of the subjects led to the large SE noted. Because each subject served as his/her own control in the ANOVA, the variability was accounted for.

 

SUMMARY

  In conclusion, this study showed that the two prototype canes and the PC decrease muscle voltage output in the upper limb during weight bearing. They do not increase muscle voltage dynamically; in fact, the PC decreases output. The wrist splint is probably not needed since this was not shown to decrease muscle output. Futher studies are underway to examine the effect of these canes on muscle voltage output.
 
ACKNOWLEDGMENTS

  The authors are indebted to Dr. Donald Marcus, MD, Chief of Rheumatology, Baylor College of Medicine, for his consultation on the grant, and to Mr. James Song, MS, for his expert assistance in biostatistical analysis of the data. We also wish to thank Noraxon for the loan of the telemetric EMG system.

 
REFERENCES
  1. Bard G, Ralston HJ. Measurement of energy expenditure during ambulation, with special reference to evaluation of assistive devices. Arch Phys Med Rehabil 1958;40:415-8.
  2. Blount WP. Don't throw away the cane. J Bone Joint Surg 1956;38A:6955-708.
  3. Fisher SV, Patterson RP. Energy cost of ambulation with crutches. Arch Phys Med Rehabil 1981;62:250-6.
  4. Khoddadadeh S. Osteoarthritic gait dynamics from force plate measurements. J Biomed Eng 1985;6:315-7.
  5. Waters RL, Perry J, Conaty P, Lunsford B, O'Meara P. The energy cost of walking with arthritis of the hip and knee. Clin Orthop 1987;214:278-84.
  6. Stevens JC, Beard CM, O'Fallon WM, Kurland LT. Conditions associated with carpal tunnel syndrome. Mayo Clin Proceed 1992;67(6):541-8.
  7. Werner R, Waring W, Davidoff G. Risk factors for median mononeuropathy of the wrist in postpoliomyelitis patients. Arch Phys Med Rehabil 1989;70(6):464-7.
  8. Waring WP III, Werner RA. Clinical management of carpal tunnel syndrome in patients with long-term sequelae of poliomyelitis. J Hand Surg Am 1989;14(5):865-9.
  9. Florack TM, Miller RJ, Pellegrini VD. The prevalence of carpal tunnel syndrome in patients with basal joint arthritis of the thumb. J Hand Surg Am 1992;17(4):624-30.
  10. Yosipovitch G, Yosipovitch Z. Acute calcific periarthritis of the hand and elbows in women: a study and review of the literature. J Rheumotol 1993;20(9):1533-8.
  11. Gray RG, Gottlieb. Hand flexor tenosynovitis in rheumatoid arthritis. Prevalence, distribution, and associated rheumatic features. Arthritis Rheum 1977;20(4):1003-8.
  12. Gelberman RH, Hergenroeder PT, Hargens AR, Lundborg GN, Akeson WH. The carpal tunnel syndrome: a study of canal pressures. J Bone Joint Surg 1981;63A(3):380-3.
  13. Kumar R, Cheng M, Scremin OU. Methods for estimating the proper length of a cane. Arch Phys Med Rehabil 1995;76(12):1173-5.
  14. Haughton JF, Little JW, Powers RK, Robinson LR, Goldstein B. M/RMS, an EMG method for quantifying upper motor neuron and function weakness. Muscle Nerve 1994;17:936-42.
  15. Diaz GY, dAverett DH, Soderverg GL. Electromyographic analysis is selected lower extremity musculature in normal subjects during ambulation with and without Protonics knee brace. J Ortho Sport Phys Ther 1997;26(6):292-8.


Go to the Contents of Vol. 41 No. 1

Back to Top